The  purity  meter  indication  varies  with  ionized
salt  concentration  and  the  temperature  of  the  coolant
flowing   through   the   cell.   The   temperature   effect   is
canceled  by  a  built-in  temperature  compensation  cir-
cuit.
The  inlet  conductivity  is  compared  to  a  preset
value  of  cell  conductance  to  actuate  an  alarm  circuit
when  the  purity  of  the  water  drops  below  the  preset
level.  In  addition,  the  purity  meter  provides  direct
readings  of  the  water  purity  at  the  inlet  and  the  outlet
of   the   demineralizer   Typical   operating   requirements
for  the  demineralizer  are  conductivity  1  micromho/cm
at   77°F   (resistivity   1   megohm/cm   at   77°F),   oxygen
content   0.1   parts   per   million   (ppm)   by   weight,   and
mechanical   filtration   0.5   microns   absolute.
When  water  has  been  circulated  through  the  sys-
tem  for  extended  periods  of  time,  a  high-resistivity  or
low-conductivity   reading   may   be   indicated   on   both
input   and   output   samples.   This   condition   is   highly
desirable  and  indicates  that  all  ionizable  material  has
been  properly  treated  and  that  the  demineralizer  is
maintaining  a  high  degree  of  purity.  When  a  system  is
filled  with  a  fresh  charge  of  water,  it  should  be  al-
lowed  to  circulate  for  approximately  2  hours  before
comparing  the  input  and  output  readings.  During  the
initial  circulation  period,  the  resistivity  readings  vary
because  of  the  mixing  action  of  water  that  has  been
treated  by  the  demineralizer  with  the  fresh  charge  of
water.
A  properly  operating  system  can  supply  water  of
acceptable  purity  in  4  to  8  hours.  Water  in  a  system
that  has  been  secured  for  any  length  of  time  should  be
of  acceptable  purity  within  2  hours.  The  resistivity  or
conductivity   reading   required   for   a   specific   installa-
tion  must  be  maintained  for  optimum  operation  of  the
cooling   water   system.
The  first  indication  of  a  problem  in  the  demineral-
izer  is  usually  indicated  by  abnormal  purity  meter
readings  (too  low/high),  an  abnormal  flowmeter  read-
ing,  and/or  alight  and  audible  warning  from  the  purity
monitor.  Some  purity  monitors  can  be  tested  for
accuracy  by  a  built-in  test  function  on  the  meter  to
establish  if  the  problem  is  in  the  purity  monitor.  If  the
purity  monitor  does  not  have  a  test  feature,  then  use
the  calibration  plug  in  place  of  one  of  the  conductivity
cells  to  test  the  operation  of  the  purity  meter.  Most  of
the  time,  only  routine  maintenance  is  required  to  re-
turn  the  demineralizer  to  its  normal  operating  condi-
tion.
Maintenance   of   the   demineralizer   consists   pri-
marily   of   the   scheduled   replacement   of   cartridges
(before   they   are   exhausted)   and   clogged   filters.
Obtaining  satisfactory  service  life  from  the  cartridges
and  filters  is  largely  dependent  on  minimizing  ex-
ternal  contamination.  Replacement  cartridges  must  be
kept  sealed  and  stored  in  a  cool,  dry  place  until  used.
The  circulating  system  must  be  kept  tight  to  reduce
the  need  for  makeup  water.  Makeup  water,  in  any
case,  should  be  as  particle-free  as  possible  and  should
not  exceed  0.065  ppm  chloride.
OXYGEN   ANALYZERS
Oxygen  analyzers  are  installed  in  some  secondary
cooling  systems  to  measure  the  amount  of  dissolved
oxygen  in  the  liquid  coolant.  The  presence  of  oxygen
causes  oxidation  that  leads  to  the  formation  of  scale  in
the  cooling  system.  An  oxygen  analyzer  has  an  oxy-
gen  sensor  installed  in  the  supply  side  of  the  second-
ary  cooling  system.
The  sensor  is  an  electrolytic  cell  in  an  electrolyte
solution  or  gel.  The  oxygen  reacts  with  the  electrolyte,
causing  a  proportional  change  in  the  amount  of  cur-
rent  flow  in  the  sensor.  The  sensors  electrical  output
is  measured  and  displayed  on  the  oxygen  analyzers
meter,  which  is  calibrated  to  read  the  oxygen  content
in  parts  per  million  or  billion.
Because  of  the  solid-state  electronics  and  the  few
components   used,   the   oxygen   analyzer   requires   very
little   maintenance   other   than   cleaning   and   changing
the  electrolyte  in  the  sensor.  When  the  meter  on  the
analyzer   requires   frequent   calibration   because   the
meter   readings   are   drifting   or   changing   sharply,   the
analyzer  has  a  bad  sensor.
2-23